Page 1 - Universal Sample Pump

Operating InstructionsUniversal Sample PumpCat. No. 224-PCXR8SKC Inc.863 Valley View RoadEighty Four, PA 15330Form 37713 Rev 1302

Page 2 - Table of Contents

8Set the Pump Period for less time than the Sample Period to initiate intermittent sampling. The pump will automatically cycle “on” and “off” at one-m

Page 3 - Description

9User Options During SamplingPause - Pause (shutdown) the pump by pressing Start/Hold. All timing data will freeze. To resume sampling, press Start/H

Page 4 - Performance Profi le

10Low Flow Applications (5 to 500 ml/min)Using Single Adjustable Low Flow Holder SetupInstall battery (see Installing the Battery Pack on page 21). Fo

Page 5 - CE marked UL Listed

11Activating the RegulatorRemove the tamper-resistant cover. Start the pump using the On/Off switch (Figure 1, #8). Press Start/Hold (Figure 1, #3). P

Page 6

12Setting or Verifying Flow RateFor a diagram of the pump, see Figure 1, page 5.Before use, allow pump to equilibrate after moving it from one tempera

Page 7 - Figure 1

13When the desired fl ow is set, place the pump in Hold by pressing Flow and Battery Check. Disconnect the calibrator. Replace the sorbent tube used fo

Page 8 - De-activating the Regulator

14SamplingBefore use, allow pump to equilibrate after moving it from one temperature extreme to another.Protect sample pump from weather when in use o

Page 9

15Low Flow Applications (5 to 500 ml/min)Using Multiple-tube Adjustable Low Flow HolderSetupFor a diagram of the pump, see Figure 1, page 5.Figure 3Qu

Page 10 - Sampling

16continued on page 17Setting or Verifying Flow RateWhen performing multiple-tube sampling using an adjustable low fl ow holder (dual, tri, or quad), e

Page 11

17Turn the fl ow adjust screw (Figure 3, #2) for the appropriate port of the low fl ow holder until the desired fl ow rate is achieved. Turn clockwise to

Page 12

Table of ContentsDescription ... 1Performance Profi le ...

Page 13 - Activating the Regulator

18Programming Delayed and Intermittent SamplingTo enter Delayed Start Mode: From Hold, press Set-up. Enter the number of minutes delay (up to 9999) be

Page 14 - Flow adjust

19SamplingBefore use, allow pump to equilibrate after moving it from one temperature extreme to another.Protect sample pump from weather when in use o

Page 15

20MaintenancePump Inlet FilterThe PCXR8 Sampler is fi tted with a fi lter/trap inside a clear plastic intake port housing. This prevents particles from

Page 16

21Installing the Battery PackTo enhance battery life, SKC ships battery packs separate from the pump. Once installed, completely charge battery pack b

Page 17 - Charger and battery

22Use of a repaired or rebuilt battery pack voids any warranty and the UL Listing for intrinsic safety.Do not charge or operate the pump from the char

Page 18 - Connect holder to

23This page is intentionally blank.

Page 19

24Parts DescriptionsUse only SKC-approved parts to ensure reliable performance. Failure to do so voids any warranty and UL Listing for intrinsic safet

Page 20

25224-PCXR8 Sample Pump See page 24 for parts listing.131415AIRCHEKSAMPLER54321SAMPLE PERIOD Ñ MINUTESSTARTHOLDFLOW AND BATTERYCHECKDIGITSETPUMPRUNTIM

Page 21

26Replacement PartsSee drawings on page 27.Pump Case PartsP21411 Case Parts (Excluding Battery Case)P21661MH Battery Pack Assembly, NiMHP21661

Page 22 - Maintenance

27Pump Case PartsPump Stack (Part #P22433RS2) ExplodedP22433RP21661MH or P21661P22433UP22433LP22433RS2(does not include fl owmeter and fi lter housing a

Page 23 - Replacing the Battery Pack

1DescriptionThe PCXR8 Universal Sample Pump is a constant fl ow air sampler suitable for a broad range of applications. It is ideal for industrial hygi

Page 24 - Pump Service

28Optional AccessoriesCalibrator Cat. No.Defender Primary Standard Calibrator, 50 to 5000 ml/min, includes lead-acid battery, charger (100-240 V), sof

Page 25

2Performance Profi leFlow Range: 1000 to 5000 ml/min (UL Listed model)(5 to 500 ml/min requires adjustable low fl ow holder)Weight: 34 oz (964 gm)Dime

Page 26 - Parts Descriptions

3Flow Fault: If the pump is unable to compensate for longer than 15 seconds due to exces-sive back pressure, the pump enters fl ow fault. During fl ow

Page 27 - 224-PCXR8 Sample Pump

4Table 2. Pump Run Time in Hours with NiMH BatteryFollowing are typical run times achieved when using a fully charged nickel-metal hydride (NiMH) batt

Page 28 - Replacement Parts

5OperationHigh Flow Applications (1000 to 5000 ml/min) SetupInstall battery (see Installing the Battery Pack on page 21). For optimum charge, ensure p

Page 29 - Pump Case Parts

6De-activating the Regulator To ensure the pump is set for high fl ow, remove the cap screw (Figure 1, #18) covering the regulator valve and turn the e

Page 30 - Optional Accessories

7Programming Delayed and Intermittent SamplingTo enter Delayed Start Mode: From Hold, press Set-up. Enter the number of minutes delay (up to 9999) bef