Developed in collaboration with the National
Institute of Working Life in Umea, Sweden
and validated by OSHA, the SKC UME
x
100
Formaldehyde Passive Sampler collects formalde-
hyde in the 5 ppb to 5 ppm range.* The UME
x
100
Passive Sampler contains a tape treated with
2,4-dinitrophenylhydrazine (DNPH). Each sampler
incorporates a “blank/correction” section of tape in addition
to the sample section of tape. Analysis is by high-performance
liquid chromatography (HPLC) for identi cation of formaldehyde
that may be present in the sample. Designed for single use, the SKC
UME
x
100 Passive Sampler is packaged in an aluminized pouch for
easy sample transport to a laboratory.
* If sampling in an atmosphere containing formalin, see www.skcinc.com/instructions/1795.pdf for
eld study information.
Application Notes
Because of the UME
x
100 Samplers relatively high sampling rate and sensitive
analysis, it can be used for 15-minute (STEL) sampling in the ppm range, personal
monitoring up to 12 hours, and 24-hour to 7-day monitoring of contaminants found
in indoor environments.
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: [email protected]
Passive Sampler for Formaldehyde
Cat. No. 500-100
www.skcinc.com
Vue de la page 0
1 2 3 4

Résumé du contenu

Page 1 - Cat. No. 500-100

Developed in collaboration with the National Institute of Working Life in Umea, Sweden and validated by OSHA, the SKC UMEx 100 Formaldehyde Passive Sa

Page 2 - Sampling Instructions

Performance Profi leSampling Rate 28.6 ml/min with a relative standard deviation of 7.6% at a windfor Formaldehyde: velocity of 5 to 100 cm/sec for 1

Page 3 - Analysis Instructions

1. Open the pouch and remove the sampler. Do not discard the pouch as it is used to send the sampler to the laboratory. Store the pouch away from po

Page 4

Form 3760 Rev 14052. The formaldehyde 2,4-dinitrophenylhydrazone is detected with UV absorp-tion between 230 and 370 nanometers (nm). Tunable UV det