BioSampler®Operating InstructionsSKC Inc.863 Valley View RoadEighty Four, PA 15330 USAForm 37084 Rev 1404
Description The patented* SKC BioSampler is a highly effi cient bioaerosol and biologically inert airborne particle collection device that traps airbo
OperationCalibrationThe BioSampler’s tangential nozzles act as sonic orifi ces. Therefore, if the pump maintains a pressure drop of 0.5 atm (15 in Hg)
Sampling with Collection LiquidThe BioSampler can be used for both short-term and long-term sampling depending on the collection liquid chosen. Oper
Caution: When operating correctly, the collection liquid in the BioSampler should be swirling gently up the inner wall of the sampler, we ing it whe
9. Accurately measure the fi nal volume of collection liquid in an aseptic area and aseptically transfer it to a sterilized container for analyses.No
Transporting Samples1. All individual sample containers should be sealed to prevent contamination during transport and properly labeled with the sam
• ImmunoassayImmunoassay is used for the quantifi cation of airborne allergens (e.g., dust mites and animal dander). It relies on the binding of antib
BioSampler Kits and Accessories Description Cat. No.BioSampler, 3-piece glass including inlet section, outlet section, and collection vessel (does not