Operating Instructions
AirChek
®
XR5000
SKC Inc.
863 Valley View Road
Eighty Four, PA 15330 USA
Form 38047 Rev 1401
Vue de la page 0
1 2 3 4 5 6 ... 21 22

Résumé du contenu

Page 1 - Operating Instructions

Operating InstructionsAirChek® XR5000SKC Inc.863 Valley View RoadEighty Four, PA 15330 USAForm 38047 Rev 1401

Page 2 - Table of Contents

7Locking the KeypadLocking: In any mode, press  5 times quickly. A flashing “L” will appear in the lower right corner of the display.Unlocking: Press

Page 3 - AirChek XR5000 Quick Guide

8User Setup FunctionsFunction OverviewUser setup functions are listed below in the order in which they display. Note that the CLr function for clearin

Page 4 - Description

9Setting a Timed RunProgram the AirChek XR5000 from its keypad to run from 1 to 9999 minutes.With pump in a non-running state (no fl ashing blue LED):1

Page 5 - Performance Profi le

10During a run delay with continuous run:a. The blue LED on top of the pump case will fl ash during the run delay even though the pump is not running.

Page 6

11f. At the end of the run, the pump will stop and “Timed Run Remaining” and “0” will display. To display accumulated run time, press and hold .g.

Page 7

12CalibrationCalibration (High Flow: 1000 to 5000 ml/min)Before use, allow pump to equilibrate after moving it from one temperature extreme to another

Page 8

13CalibrationCalibration (Low Flow: 5 to 500 ml/min)Requires Constant Pressure Controller (CPC) and Adjustable Low Flow Tube Holder - see Accessories,

Page 9 - Turning the Pump Off

14Sampling• Before use, allow pump to equilibrate after moving it from one temperature extreme to another.• Use of any device or battery pack other

Page 10 - Checking Battery Charge Level

15NoteTipA low battery fault may occur instead of a fl ow fault when there is a low battery charge at the time of the fault, excessive back pressure, a

Page 11 - User Setup Functions

16Battery ReplacementNote122STOP! See www.skcinc.com/instructions/1756.pdf for more information on SKC pump battery packs.Replacing the Li-Ion Batter

Page 12 - Setting a Timed Run

Description ...1Performance Profi l

Page 13

17Resetting Pump to Manufacturer SettingsIf the pump does not operate as expected, perform the following procedure:1. Remove the battery pack (see Re

Page 14

18AccessoriesDescription Cat. No.Defender Primary Standard Calibrator, 50 to 5000 ml/min fl ow range, includes lead-acid battery, charger (100-240 V)

Page 15 - Calibration

19Description Cat. No.Battery Packs 4-cell Li-Ion* Battery Pack P85004 2-cell Li-Ion* Battery Pack P85002Accessories Belt Clip P20139 Filter/O-

Page 16

AirChek XR5000 Quick GuideKeypad Basics (star key) Scrolls through parameters in user setup functions. (up/down arrow keys) Increase or decrease

Page 17 - Sampling

1DescriptionSKC AirChek XR5000 Sample Pumps are designed to offer users enhanced battery power and easy operation in a lightweight pump that provides

Page 18

2Flow Range: 1000 to 5000 ml/min (5 to 500 ml/min requires optional low fl ow adapter kit)Compensation Range: 5000 ml/min at 10 inches water back pre

Page 19 - Battery Replacement

3(model dependent)Performance Profi leAuto-off: 5 minutes of inactivityBattery Pack: High-power Li-Ion (4 cell), rechargeable, 7.4 V, 4.4-Ah capacity

Page 20 - Maintenance

4Charging the Lithium-Ion Battery PackFor models containing a lithium-ion battery pack only.Shown with single charger (Cat. No. 222-241). A fi ve-stati

Page 21 - Accessories

5Power supply jackCharge status LED See http://www.skcinc.com/instructions/1756.pdf for more information on SKC pump battery packs.Reading the Chargi

Page 22 - Replacement Parts

6SetupKeypad BasicsThe AirChek XR5000 operates by pressing key sequences on the keypad located on the front of the pump case.Keys Scrolls through pa